6000V漏电起痕试验仪 制造标准:
IEC60112:2009、GB/T4207-2012、UL746A、ASTM D 3638-2007、DIN53480等标准设计进行制造的。
6000V漏电起痕试验仪 用途
漏电起痕试验仪。适用于固体绝缘材料在潮湿条件下相比漏电起痕指数和耐漏电起痕指数的测定方法,具有简便、准确、可靠、实用等特点。
技术指标
1.电极尺寸:5mm?0.1mm*2mm?0.1mm,电极一端边缘切成30度角的斜面;
2.电极材料:铜 铂金(铂金长度:≥12mm,纯度≥99.0%);
3.电极间相距:4.0mm?0.1mm;
4.每个电极对试样作用力:1.0N?0.05N;
5.试样品尺寸:≥15*15mm, 厚度:≥3mm;
6.施加电压:100~600V(48~60Hz)之间可调,短路电流在1.0A?0.1A时电压下降不超过10%;
7.判断回路:短路电流大于0.5A时间维持2秒钟继电器动作,切断电流,指示试品不合格;
8.溶液滴落间隔时间:30s?5s;
9.滴液针头到样品表面高度:30~40mm;
10.控制液滴采用电磁阀.液滴大小:(20 30)mm3,50滴为(0.997~1.147g), 20滴为(0.380-0.480g);
11.滴液装置:试验前需要调整;
12.针头外径:0.9~1.1mm;
13.箱体内部容积:0.5m3(可选购0.75m3或1m3);
14.外部尺寸: 宽1170mm*深630mm*高1330mm;
15.箱体材料:铁板喷涂;
16.排气孔:?100mm;
17.工作电源:220V,10A,50Hz。
配件清单:
序号 | 名称 | 数量 |
1 | 电子遥控器 | 1个 |
2 | 玻璃方块 | 1块 |
3 | 铂金电极 | 1对 |
4 | 试验针头 | 3个 |
5 | 不锈钢量规4mm | 1个 |
6 | 电源线 | 1根 |
7 | 保险丝 | 2个 |
8 | NH4CL(氯化铵) | 1瓶 |
9 | 掌式电子称(校准电极压力) | 1个 |
10 | 排烟软管(可伸至3.5m) | 1根 |
11 | 升降台 | 1个 |
主要配件
序号 | 型号 |
1 | 箱体(可选不锈钢箱体) |
2 | 变压器 |
3 | 调压器 |
4 | 继电器及底座 |
5 | 漏电保护器 |
6 | 按钮 |
7 | 计时器 |
8 | 短路电流智能表 |
9 | 温控器 |
10 | 导线 |
11 | 计数器 |
12 | 无线控制器 |
13 | 电磁阀 |
Manufacturing standards:
Designed and manufactured according to standards such as IEC60112:2009, GB/T4207-2012, UL746A, ASTM D 3638-2007, DIN53480, etc.
purpose
Leakage trace tester. This method is suitable for measuring the leakage marking index and leakage resistance marking index of solid insulation materials under humid conditions, and has the characteristics of simplicity, accuracy, reliability, and practicality.
Technical indicators
1. Electrode size: 5mm?0.1mm * 2mm?0.1mm, with one end of the electrode edge cut into a 30 degree slope;
2. Electrode material: copper platinum (platinum length:≥12mm, purity≥99.0%);
3. Distance between electrodes: 4.0mm?0.1mm;
4. Each electrode exerts a force on the sample: 1.0N?0.05N;
5. Sample size:≥15 * 15mm, thickness:≥3mm;
6. Applied voltage: adjustable between 100-600V (48-60Hz), with a voltage drop of no more than 10% when the short-circuit current is 1.0A?0.1A;
7. Determine the circuit: If the short-circuit current is greater than 0.5A, maintain the relay action for 2 seconds, cut off the current, and indicate that the test sample is unqualified;
8. Dropping interval of solution: 30s?5s;
9. Dropping needle to sample surface height: 30-40mm;
10. Control the droplets using an electromagnetic valve. Droplet size: (20 30) mm3, 50 droplets are (0.997~1.147g), and 20 droplets are (0.380~0.480g);
11. Dropping device: needs to be adjusted before the experiment;
12. Needle outer diameter: 0.9~1.1mm;
13. Internal volume of the box: 0.5m3 (optional 0.75m3 or 1m3);
14. External dimensions: width 1170mm * depth 630mm * height 1330mm;
15. Box material: Iron plate spray coating;
16. Exhaust hole: ? 100mm;
17. Working power supply: 220V, 10A, 50Hz.
List of accessories:
Number Name Quantity
1 electronic remote control
2 glass blocks, 1 piece
3 platinum electrodes, 1 pair
4 test needles, 3 pieces
5 stainless steel gauges, 4mm, 1 piece
6 power cables, 1 piece
7 fuses, 2 pieces
1 bottle of 8 NH4CL (ammonium chloride)
9 handheld electronic scales (calibrated electrode pressure) 1
10 smoke exhaust hoses (can extend up to 3.5m) 1 piece
11. One lifting platform
Main accessories
Serial Number Model
1 box (optional stainless steel box)
2 Transformers
3 Voltage regulator
4 Relay and base
5 Leakage protectors
6 buttons
7 Timer
8 Short circuit current intelligent meter
9 Temperature controller
10 wires
11 Counters
12 wireless controllers
13 solenoid valve
业务咨询:932174181 媒体合作:2279387437 24小时服务热线:15136468001 盘古机械网 - 全面、科学的机械行业免费发布信息网站 Copyright 2017 PGJXO.COM 豫ICP备12019803号